5.4.5
- Anti-Crush Sensor
Installed at the end of the leg rest, the mechanical pressure sensor (1)
acts to prevent accidents caused by the presence of any obstacle in the
movement of the leg rest.
When the anti-crush sensor is activated, the equipment emits a
continuous beep and locks all equipment commands, except the
Seat Up
command. Raise the seat to deactivate the sensor and then check and
remove the obstacle in the area that activated the sensor.
5.4.6
- Operational Warning Sounds
The Performance Table beeps to identify some operations. Check out the warning sounds and their
applications in the table below:
Aviso Sonoro
Aplicação
1 long beep
- When turning on the equipment
1 short beep
- When selecting a work position
2 short beeps
- When selecting or recording a work position
- When pressing the Zero Position command
- When reaching the light intensity limits (minimum and maximum) of the operating light
Premium LED
3 short beeps
- Error command*
- End of the interval for work position selection or recording
- End of positioning of a working position and Zero Position when the seat is already in
the Zero Position
4 short beeps
- When cancelling a continuous or automatic movement
2 long beeps
- Interval for setting up a working position or switching operating mode
1 continuous beep
- The anti-crush sensor is activated
*Error command: command that the equipment cannot perform; for example, press the command
Seat Up
when the seat is already in the highest position.
5.4.7
- Positioning of the Patient and the Operator
During the equipment’s movement, the operator and other persons close to
him/her should position
themselves on the equipment sides, respecting the minimum distance of 50 centimeters, remaining out of
the movement area of the equipment and its components.
To ensure safe and proper patient positioning, the operator should advise him to keep his hands at
sight during the entire equipment movement, on the equipment upholstery or on the specific accessories for
performing the procedure. The operator must advise the patient to remain in this position during the whole
equipment movement. If the operator sees any situation that may generate any kind of risk to the patient,
the chair must be stopped immediately.
The presence of the operator, persons or objects in the movement areas of the equipment and its
components may cause damage to the equipment and / or impair its correct operation.
All equipment’s automatic movements must be supervised by the operator. If the operator sees any
situation that may generate any kind of risk to the patient, the chair must be stopped immediately.
5.4.8
- Emergency Position
Press the Backrest Down command (
section 5.4
) until the end of the course. The equipment will adjust
the position of the backrest by -5º (for Performance Table with 3 motors) in relation to the floor. If the
equipment has an Incline Seat command, press it by tilting the seat for back to the end of the course to
adjust the incline by -10°.
Summary of Contents for PERFORMANCE TABLE
Page 1: ......
Page 2: ......
Page 24: ...24 11 Dimensional...
Page 25: ...25 12 Simbology According to IEC 60601 1 and IEC 60878 standards...
Page 35: ...35...
Page 37: ......
Page 38: ......
Page 60: ...24 11 Dimensional...
Page 61: ...25 12 Simbologia La simbolog a est en conformidad con las normas IEC 60601 1 y IEC 60878...
Page 71: ...35...