7.4 Remove and Re-Install the Convection Channel
1.
Open the Sliding Top Cover.
2.
Switch off the CERO.
3.
Remove the Front Cover (see chapter 7.3).
4.
Gently press the Convection Channel clips at the top and bottom to free it from the
chamber (figure 46).
Figure 46: Remove the Convection Cha nnel
To re-install the Convection Channel back in place, proceed as follows:
1.
Position the Convection Channel.
2.
Press it gently in place until you hear the clips snap.
7.5 Replace Hepa Filters
1.
Remove the 3 Hepa Filters by pulling them out of their position (figure 47).
2.
Take 3 new Hepa Filters and push them in place.