CERO_Operators Guide_Version 2018-12_01
47
6.4.2.2 Set Notification Level
The CERO notifies when the pH is becoming too low, i.e. lower than the defined notification
level. Please follow the instructions below to set the notification level:
1. In the System Menu, go to Settings.
2. Enter the Reader Window (figure 35).
3. Adjust the notification level of the quantitative pH Measurement by pressing the Minimal
pH Button.
4. Adjust the notification level of the qualitative color measurement by pressing the Minimal
Color Button.
5. Rotate the click-wheel to set the notification level.
6.4.3 Enable the pH or the Color Mode
1. In the System Menu, go to Settings.
2. Enter the Reader Window and select pH Settings (figure 35).
3. Enable the desired Monitoring Mode (figure 36). Two monitoring modes can be enabled: a
qualitative color and a semi-quantitative pH mode (see chapter 6.2).
Figure 36: pH Settings (Color Monitoring Mode)