The only means of checking that the instrument’s detection function is working
properly is the
calibration
performed with the help of a
standard gas
.
Calibration must be done :
-
At least once a year following the use and the type of sensor
-
whenever the instrument has been exposed to overly high concentrations of
gas
-
when an instrument has been stored for more than one month and without
use
-
when changing of sensors
-
when fault appears
Within the calibration the menu "visible and audible" alarms are not triggered.
3.1.
Calibration procedure
“ Calibration ” menu consists of
-
setting the zero in clean air and the sensitivity with calibration gas in
) version (toxic gases).
-
Adjust oxygen sensitivity in OX 2000(+) version.
3.2.
Adjusting the zero in clean air (only for toximeter version)
Display the “ programming ” menu
To display the "calibration" menu
Display the"calibration" menu
To enter the “ calibration ”menu
P
r
G
CAL
ENTER