19
DISPLAY
CAUSE
REMEDY
BAT
pictogramme
Dry batteries are worn :
the instrument is no
longer usable.
-
Switch off the instrument
-
Alarm cannot be cleared
-
Replace dry batteries pack
OR
Measurement range
exceeded
-
Alarm can be cleared
manually
Zero fault : sensor zero
too much shifted to low
or high
- Calibrate
Sensibility fault : low
sensitivity
-
Recalibrate and if the fault
persists : the sensor must be
replaced.
Worn out sensor fault :
low sensitivity
-
alarm cannot be cleared :
-
Replace sensor and calibrate
again
Eeprom fault : eeprom
memory is not initialized
-
Return the instrument into
factory to be reinitialised.
Measurement too
negative or sensor faulty
-
Recalibrate and if the fault
persists : the sensor must be
replaced
Problem of
communication between
micro and eeprom
-
Return the instrument into
factory to be repaired
Component fault : T°
detector is faulty
-
Return the instrument into
factory to be repaired
Wrong type of sensor.
Alarm cannot be cleared
Connect the correct type of
sensor
Sensor absent
-
Alarm cannot be cleared : fit
the sensor after switching
off the instrument
(cyclic)
(cyclic)
(cyclic)
(cyclic)
DEF
1
DEF
2
(cyclic)
(cyclic)
DEF
4
(cyclic)
DEF
8
DEF
16
ERR
35
69
DEF
64
(cyclic)
à
DEF
128
DEF
128