26
Figure 32. Cycle Start Time
At set time the cycle will start. A thermal cycling icon will be displayed on the Homepage as a reminder.
Trying to change set points or starting a new thermal cycle while a different one is still running will open
the window in Figure 33
. You can press Abort Cycle to modify it or start a new one. Pressing ‘View’ will
open the current thermal cycle settings.
Figure 33. Running Cycle window
6.2.1.5
Open Incubator
During the normal operation it could be necessary to open the incubator. The picture below shows the
procedure to follow when opening incubator:
1.
Press on
Open Incubator
, before removing the lid. Now you can open the incubator;
2.
Once you are done close the chamber and wait for 2 minutes in order to let the environment achieve
steady state; press on
Incubator Closed
and the temperature controller will resume operating
normally.
Figure 34. Open incubator
1
2