background image

711449 (Rev.3)  02/2016

Appendix D
Troubleshooting

Table D-1. Troubleshooting

PROBLEM

CAUSE

CORRECTIVE ACTION

Display is black

Monitor is turned OFF or Battery is 
expired.

Press 

I/O

 to turn ON monitor

If monitor does not respond, replace the battery
If problem persists, call Ohio Medical Corporation service

No response to keypad 
commands

Battery is expired

Press 

I/O

 to turn ON monitor

If monitor does not respond, replace the battery
If problem persists, call Ohio Medical Corporation service

         displays and 
visual and audible 
alarms activate

Cable is disconnected or malfunctioning Check cable and sensor connections

Sensor is disconnected

Visually check cable for damage. If necessary, replace 
cable

Sensor membrane is perforated

If alert persists, replace sensor

The thermistor circuit is open on the 
sensor circuit board

If alert persists following sensor replacement, call Ohio 
Medical Corporation service

During calibration, “CAL 

(55´ÀDVKHVYLVXDO

and audible alarms 
activate; then, “CAL” 

ÀDVKHV

Calibration performed using a gas other 
than room air or 100% oxygen or wrong 
calibration key pressed (100%/21%)

Recalibrate the monitor, ensure proper calibration key 
pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio 
Medical Corporation service

During operation, “CAL” 

ÀDVKHV

Sensor signal is out of monitoring range

Recalibrate the monitor, ensure proper calibration key 
pressed for calibration gas of room air or 100% oxygen
If alert persists, replace sensor
If alert persists following sensor replacement, call Ohio 
Medical Corporation service

           displays and 
monitor beeps every 30 
seconds

Monitor has approximately six hours of 
operating time remaining

Replace the battery as soon as possible; then recalibrate 
the monitor and reset the high and low alarm values

GLVSOD\V³´ÀDVK

-

es, both visual alarms 
activate, and monitor 
emits a three-tone beep 
for four minutes; moni-
tor shuts OFF

Battery is expired and monitor is not 
operating

Replace the battery as soon as possible; then, recalibrate 
the monitor and rest the high and low alarm values.

“ERR” and error code 
(01 to 08) display. Vi-
sual and audible alarms 
activate; monitor locks

Microprocessor detects internal error

Note error code and see 

Appendix E, Error Codes

Disconnect battery

Call Ohio Medical Corporation service

Possible physical    
damage to instrument

Instrument is dropped and/or exposed 

WRÀXLGV

Turn instrument OFF, then ON to initiate self-test.  If 
instrument is damaged, “ERR” and error code (01 to 08) 
display. Disconnect battery. Call Ohio Medical Corporation 
service

Appendix E 

MiniOX

®

 3000 Oxygen Monitor   

 

 

 

 

                 Appendix D, Troubleshooting

D-1

Summary of Contents for MiniOX 3000

Page 1: ...T l fax 1 608 222 6028 Manuale d istruzioni Assistenza al cliente 1 847 855 0800 Manutenzione 1 847 855 0800 Fax 1 608 222 6028 FEDERAL U S LAW RESTRICTS THIS PRODUCT TO SALE BY OR ON THE ORDER OF A P...

Page 2: ...HQ LGLRP DV 3DUD HQFRQWUDU OD VHFFLRQ HVF ULWD HQ HO LGLRPD GH VX SUHIHUHQFLD SRU IDYRU HOLMD HO VHSDUDGRU FRQ HO FRORU LQGLFDGR D FRQWLQXDFLRQ H PDQXHO FRQWLHQW GHV LQVWUXFWLRQV HQ ODQJXHV 3RXU WURX...

Page 3: ...o these products are voided if the products are not installed used and serviced in accordance with the instructions in this manual Please protect yourself and your patients by following them We encour...

Page 4: ...ng warranties shall not apply if the product has been repaired other than by Ohio Medical or in accordance with written instructions provided by Ohio Medical or altered by anyone other than Ohio Medic...

Page 5: ...the unit OFF and contact your nearest Ohio Medical Service Center for additional information 8 Use of devices generating or emitting electromagnetic radiation near the MiniOX 3000 Oxygen Monitor may i...

Page 6: ...MiniOX 3000 Oxygen Monitor 2 1 Figure 2 1 MiniOX 3000 Oxygen Monitor 2 1 CAUTION 2 2 Figure 2 2 Assembled MiniOX 3000 Oxygen Monitor 2 2 Section 3 Operation 3 1 Calibration 3 1 CAUTION 3 1 To Calibrat...

Page 7: ...sor Replacement 4 2 H HFWRU 5HSODFHPHQW 4 2 Cable Replacement 4 2 Cleaning Disinfection and Sterilization 4 2 CAUTION 4 2 Instrument 4 2 Sensor and Cable 4 3 6HQVRU H HFWRU 5HWDLQLQJ 6WUDS DQG 7HH GDS...

Page 8: ...ases and Vapors C 1 Table C 1 Interferent Gases and Vapors C 1 Appendix D Troubleshooting D 1 Table D 1 Troubleshooting D 1 Appendix E Error Codes E 1 Table E 1 Error Codes E 1 Appendix F Accessories...

Page 9: ...rm and 18 to 99 low alarm or the default high low settings may be used 50 and 18 respectively Audible and visual alarms activate when oxygen concentrations fall below the preset or default low alarm s...

Page 10: ...ical current displaying O2 values The Ohio Medical Corporation oxygen sensor is self zeroing when no oxygen is present no current is produced 0 oxygen is displayed The sensor has a minimal response to...

Page 11: ...using a calibration gas other than room air or 100 oxygen or if the microprocessor detects a calibration error during normal operation WKH GLVSOD DVKHV D ZDUQLQJ PHVVDJH 1 3 each time the monitor is...

Page 12: ...ons a test function that allows the operator to ensure that the high and low alarms activate at the preset alarm levels See Section 3 Test Functions Radio Frequency Electromagnetic Insensitivity The M...

Page 13: ...ration Manual CD 2 Check the sensor s manufacturing date on the sensor package If the sensor is put into service within six months after this date the sensor will PHHW SHUIRUPDQFH VSHFL FDWLRQV 3 Remo...

Page 14: ...sor unnecessarily during calibration or use Body heat can cause the sensor s thermistor to change disproportionately to the change in gas sample temperature at the sensing electrode This can produce s...

Page 15: ...WUDWLRQ RI When a more precise measurement is desired repeat calibration using 100 oxygen concentration Changes in pressure humidity or temperature may affect calibration accuracy See Appendix B Effec...

Page 16: ...at 100 O2 Figure 3 2 Note Prior to calibrating at 100 O2 the MiniOX 2 JHQ 0RQLWRU PXVW UVW EH FDOLEUDWHG in room air 1 Calibrate in room air See Section 3 To Calibrate In Room Air 2 Expose the sensor...

Page 17: ...ibrate in room air 2 Place the sensor in a stream of oxygen with a known concentration of 100 until the reading stabilizes 3 Recalibrate the monitor 4 After calibration measure room air This reading s...

Page 18: ...eactivates If the alarm condition is corrected and then reoccurs within this interval both audible and visual alarms activate 2 To manually terminate the silence mode press the down DUURZ ODEHOHG ZLWK...

Page 19: ...Medical Corporation service During calibration CAL ERR DVKHV YLVXDO DQG DXGLEOH DODUPV DFWLYDWH WKHQ DVKHV Calibration performed using a gas other than room air or 100 oxygen or wrong calibration key...

Page 20: ...m condition Installing the Sensor in a Breathing Circuit To use the sensor in a breathing circuit you will need the following 6HQVRU ZLWK GH HFWRU Tee adapter Retaining strap To install the sensor in...

Page 21: ...711449 Rev 3 02 2016 SENSOR CABLE RETAINING STRAP SENSOR GASKET DEFLECTOR O RING 15MM TEE Figure 3 3 Installing Sensor into Tee Adapter MiniOX 3000 Oxygen Monitor Section 3 Operation 3 7...

Page 22: ...re 3 4 Sensor in Breathing Circuit FASTEN RETAINING STRAP TO POSTS ON TEE 15MM TEE ADAPTER DEFLECTOR SENSOR SENSOR CABLE RETAINING STRAP SLIDE DOVETAIL COMPLETELY INTO BRACKET MOUNTING BRACKET INSTALL...

Page 23: ...battery cover is properly VHDWHG DQG DW RQ WKH EDFN RI WKH 0LQL2 3000 Oxygen Monitor case 7 Recalibrate the monitor Reset the low and high alarms if desired Section 4 Maintenance and Care Use only gen...

Page 24: ...RU assembly Cable Replacement To replace the coiled cable 1 Verify that the monitor is turned OFF The display should be blank 2 Disconnect the old cable from the sensor 3 Disconnect the old cable from...

Page 25: ...are used Sterilization 7KH VHQVRU GH HFWRU UHWDLQLQJ VWUDS DQG WHH DGDSWHU may be sterilized using Cidex per manufacturer s instructions steam or ethylene oxide Due to the varying conditions imposed...

Page 26: ...n sealed package Battery Life Approximately 1500 device hours assuming 80 ON 20 OFF Operating Temperature Range 0o C to 40o C 32o F to 104o F Storage Temperature 40o C to 70o C 40o F to 158o F Humidit...

Page 27: ...9ROWDJH XFWXDWLRQV LFNHU HPLVVLRQV IEC 61000 3 3 Not applicable Not applicable Guidance and manufacturer s declaration electromagnetic immunity The MiniOX LV LQWHQGHG IRU XVH LQ WKH HOHFWURPDJQHWLF H...

Page 28: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency...

Page 29: ...nce by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the MiniOX 3000 as recommended below according to the maximum output power of the communi...

Page 30: ...creases the actual concentration of oxygen Humidity in a sample has the same effect as diluting the sample with another gas For example if 100 oxygen is satu rated with 100 humidity the actual concent...

Page 31: ...NT VOLUME DRY INTERFERENCE EQUIVALENT OF O2 Carbon Dioxide 10 0 1 Cyclopropane 50 0 2 HV XUDQH 7 5 0 1 Diethyl Ether 20 0 3 Q XUDQH 4 0 1 Halothane 5 0 2 Helium 80 0 2 VR XUDQH 3 0 5 0HWKR XUDQH 4 0 1...

Page 32: ...sensor If alert persists following sensor replacement call Ohio Medical Corporation service During operation CAL DVKHV Sensor signal is out of monitoring range Recalibrate the monitor ensure proper ca...

Page 33: ...es not reoccur press set 3 The battery is disconnected If an error reoccurs the monitor is inoperative and must not be used call Ohio Medical Corporation Repair at 1 847 855 0800 The error code that a...

Page 34: ...ration Manual CD 600700 Operating Manual printed 711449 MiniOX 3000 Oxygen Monitor 711451 Battery Cover Assembly 710974 Monitor Stand 655033 H HFWRU 803229 Dovetail 474606 Screw 4x1 4 set of 4 thread...

Reviews: