711449 (Rev.3) 02/2016
Sensor and Cable
Cleaning
The external surfaces of the oxygen sensor and of
the cable may be cleaned by wiping them with a cloth
moistened with a mild detergent solution.
Disinfection
The external surfaces of the oxygen sensor housing
and of the cable may be disinfected by wiping them
with a cloth moistened with ethanol or Cidex. The
instrument is not designed to withstand the conditions
imposed by steam, ethylene oxide or radiation
sterilization.
6HQVRU'HÀHFWRU5HWDLQLQJ6WUDSDQG7HH
Adapter
Cleaning
7KHVHQVRUGHÀHFWRUUHWDLQLQJVWUDSDQGWHH
adapter may be cleaned by wiping them with a cloth
moistened with a mild detergent solution. The parts
must be thoroughly dry before they are used.
Disinfection
7KHVHQVRUGHÀHFWRUUHWDLQLQJVWUDSDQGWHHDGDSWHU
may be disinfected by washing them with ethanol or
Cidex (per manufacturer’s instructions). The parts
must be thoroughly dry before they are used.
Sterilization
7KHVHQVRUGHÀHFWRUUHWDLQLQJVWUDSDQGWHHDGDSWHU
may be sterilized using Cidex (per manufacturer’s
instructions), steam or ethylene oxide. Due to the
varying conditions imposed on materials during
sterilization, it is not possible to determine the exact
number of times sterilization processes can be carried
out. Therefore the operator must carefully examine
WKHVHQVRUGHÀHFWRUUHWDLQLQJVWUDSDQGWHHDGDSWHU
after sterilization and prior to use to verify that the
LWHPLV¿WIRUXVH7KHRSHUDWRUPXVWYHULI\WKDWWKH
items are free from tears and cracks and that the
items have not undergone any material changes that
PD\FRPSURPLVHWKHLU¿WQHVVIRUXVHHJEULWWOHQHVV
and dimensional changes). The operator must also
examine the items to verify that the items are free
of chemical residuals resulting from the sterilization
process.
Because of the variability of cleaning, disinfection
and sterilization processes, Ohio Medical Corporation
FDQQRWSURYLGHVSHFL¿FVWHULOL]DWLRQLQVWUXFWLRQVQRU
can the sterility of an item be ensured.
MiniOX
®
3000 Oxygen Monitor
Section 4, Maintenance and Care
4-3