MiniOX
®
3000 Oxygen Monitor
Section 3, Operation
• “CAL”
•
• “21% CAL”
4. Press . The following appears:
• “CAL”
• “21% CAL”
• a 10-segment bar graph that “counts down”
two seconds per bar for 20 seconds
5. After 20 seconds, the calibration process is
complete the device:
• displays 20.8% ±2%O
2
(18.8% to 22.8%)
• proceeds to the monitoring mode
• displays the current oxygen concentration
as %O
2
To Calibrate at 100% O
2
(Figure 3-2):
Note: Prior to calibrating at 100% O
2
, the MiniOX
®
2[\JHQ0RQLWRUPXVW¿UVWEHFDOLEUDWHG
in room air.
1. Calibrate in room air (
See Section 3, To
Calibrate In Room Air)
.
2. Expose the sensor to 100% oxygen and allow
the readings to stabilize prior to initiating the
calibration.
•
³&$/´ÀDVKHVLQWKHGLVSOD\
3. Press 100%. The following appears on the
display:
• “CAL”
•
• “100% CAL”
4. Press The following appears on the display:
• “CAL”
• “100% CAL”
• a 10-segment bar graph that “counts down”
two seconds per bar for 20 seconds
5. After 20 seconds, the calibration process is
complete; the device:
• displays 100.0% +0/-2% (98% to 100%)
• proceeds to the monitoring mode
• displays the current oxygen concentration
as %O
2
3-2
Figure 3-2
Calibrating the MiniOX
®
3000 Oxygen Monitor at 100% O
2