711449 (Rev.3) 02/2016
Battery & Battery Cover Replacement
The MiniOX
®
3000 Oxygen Monitor requires one
standard 9-volt alkaline battery. To replace the battery:
1. Verify that the monitor is turned OFF. The display
should be blank.
2. Pull out the support stand from the back of the
case.
3. Unscrew the two screws on the battery cover in
the back of the instrument and remove cover.
See Figure 4-1.
4. Remove the battery from the case and unsnap the
battery from the battery holder.
NOTE: To ensure proper start-up, wait at least
45 seconds before connecting the
fresh battery to the battery connector.
5. Snap the terminal of the new battery into the
battery holder.
6. Install the battery cover and screw into place.
Make sure that the battery cover is properly
VHDWHGDQGÀDWRQWKHEDFNRIWKH0LQL2;
®
3000
Oxygen Monitor case.
7. Recalibrate the monitor. Reset the low and high
alarms, if desired.
Section 4
Maintenance and Care
Use only genuine Ohio Medical Corporation
replacement parts when performing any
maintenance procedures included in this
manual. Failure to do so may seriously
impair the monitor’s performance.
Repair or alteration of the MiniOX
®
3000
Oxygen Monitor beyond the scope of the
maintenance instructions or by anyone other
than an authorized Ohio Medical Corporation
service person could cause the product to
fail to perform as designed
.
This section describes the following general
maintenance and care procedures for the MiniOX
®
3000 Oxygen Monitor:
• Battery Replacement
• Sensor Replacement
•
'HÀHFWRU5HSODFHPHQW
• Cable Replacement
• Cleaning
Figure 4-1
Battery Cover Removal
MiniOX
®
3000 Oxygen Monitor
Section 4, Maintenance and Care
4-1