APPENDIX B THE SERVICE MENU
B.4 Span
Calibration
Span calibration from the service menu allows you to set a new zero and maximum setting.
This is distinct from user level span calibration, which allows a user to adjust the zero and
maximum setting within the range established by the service menu span setting.
Span calibration automatically follows Linear calibration. To start from the Service Menu, press
and hold
/
Cal
and
UNIT
together.
RAMP
appears. Press
No
to bypass Ramp.
Lin
appears. Press
No
.
Filter
appears. Press
No
. (This adjusts the speed with which the scale
stabilizes. The factory setting is
High
, and should not normally be changed.)
SPAN
appears. Press
Yes
.
SPAn
The display shows
-C-
while the scale acquires the
zero value.
-C-
The specified maximum mass weight flashes.
Place the indicated weight in the center of the Pan.
200
Pressing
Yes
after placing the prompted weight accepts
the span weight shown on the display. The display then
shows
-C-
while the scale acquires the span value.
-C-
After the span value is acquired, the display shows
done
for 2 seconds and returns to
SPAN
. Press
No.
Lft.rES
appears. Press
No
.
End
appears. Press
Yes
.
The scale returns to normal weighing mode.
done
NOTE:
If calibration fails, ensure that the test area is free from drafts and the surface the scale
rests on is level and free of vibrations. Then try to calibrate again. If it continues to fail, there
may be an internal problem. To resolve internal problems, follow procedures in Chapter 3.
Traveler™ Series Service Manual
B-3
Ohaus Corporation www.ohaus.com