STARTER 5000 Bench pH Meter EN-7
Information, measure and calibrate tap area
The information area is also the shortcut to enter the setting interface:
"
Endpoint:"
Tap this area to enter the interface to set the endpoint mode. Available
modes are Auto, Manual and Time endpoint. The current endpoint mode is
displayed.
"
Sensor ID:"
Tap this area to enter the interface to select, set or change the sensor ID.
The current sensor ID is displayed.
"
Sensor Cond:"
Tap this area to enter the sensor calibration data interface. The current
sensor performance icon (judged by calibration data) is displayed.
There are 3 sensor condition icons:
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is good
Slope: 90-95%
or offset: ± (15-35) mV
Electrode condition is
acceptable
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not
good or needs cleaning
Calibration reminder icon. Set the time you want the meter to alert you to do
calibration from last calibration. When the time is reached, this icon will be displayed
and blink to remind that it is time to perform a calibration.
“
Buffers:
”
This area displays the current buffer group, tap this area to enter the interface to
select the buffer group.
Start icon. Tap this icon and the meter will begin measurement, the signal
icon
will blink. The start icon area will show the stop icon
Calibration icon. Tap this icon to start calibration.
Save icon. After saving the current measurement result (endpoint reached), this
icon color will change from black to light gray
, indicating that the reading has
already been saved and cannot be saved twice.
Output icon. Output data to external device such as a printer.
If “Shut off connection” is chosen in
System Setting
►Output,
the save
icon
will be displayed .Otherwise the output icon will be displayed in the
same position.