EN-18 STARTER 5000 Bench pH Meter
7
TECHNICAL DATA
7.1
Specifications
The technical data is valid under the following ambient conditions:
Indoor use only
Altitude: Up to 2000 m
Specified Temperature range: 5°C to 40°C
Humidity: maximum relative humidity 80 % for temperatures up to 30°C decreasing linearly to
50% relative humidity at 40°C
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category II
Pollution degree: 2
Operability is assured at ambient temperatures between 5°C to 40°C
Model
STARTER 5000
Measuring range
-2.000
…20.000 pH
–2000.00…2000.00 mV
-30°C
…130°C
Resolution
0.1/0.01/0.001 pH
1/0.1/0.01 mV
0.1°C
Error limits
± 0.002 pH
± 0.03% FS mV
± 0.1°C
Calibration
1-9 points
8 predefined buffer group+1 self-defined buffer group
Memory
1000 measurements
10*10 calibration data
Power supply
100-240V~0.3A input, 9V 1.12A output AC Adapter
Size/weight
Approximately 210 x 170 x 80 mm / 0.65 kg
Ship dimension/weight
Approximately 372 x 310 x 185 mm / 2.30 kg
Display
Touch screen liquid crystal with backlight
Input
BNC, impedance > 3 x 10e+12
Ω
Cinch, NTC 30 k
Ω
Temperature-compensation
ATC & MTC
Housing
ABS