EN-16 STARTER 5000 Bench pH Meter
5.3
Calibration Setting
5.3.1
Calibration Setting
►Buffer Group
Before calibration, please make sure the current buffer group is correct for your buffer solutions.
User can select from the 8 predefined buffer groups.
User can also set self-defined buffer groups. The adjacent 2 calibration point difference should be
larger than 0.99pH.
5.3.2
Calibration Setting
►Calibration Reminder
Enter this setting to set the time, from the last calibration, after how many hours the meter should
remind to perform calibration again. The meter will show the calibration reminder icon
and
blink, to remind user to perform calibration.
5.4
Data Log
Tap Data to get into the Data log interface, including measurement data and calibration data.
You can delete all the measurement data here. The only way to delete calibration data is to
restore factory setup.
The data can be output to a USB Flash Drive (
≤4GB).