SCOUT
TM
SJX SERIES BALANCES EN-11
Span
[
SpaN
]
Initiates a span calibration procedure (zero and span)
Lin
[
LIN
]
Initiates a linearity calibration procedure (zero, mid-point and span).
End Cal
[
End
]
Advance to the next menu or return to the top of the current menu.
4.3 Setup Menu
Enter this menu to set balance parameters.
Note:
Bold always represents factory default value.
Reset
[
reset
]
Reset the Setup menu to factory defaults.
NO
= not reset
YES = reset
Filter
[
FiltEr
]
Set the amount of signal filtering.
LOW
= less stability, faster stabilization time
MED
= normal stability, stabilization time
HI
= greater stability, slower stabilization time
AZT
[
AZt
]
Set the automatic zero tracking functionality.
OFF
= disabled
0.5d
= the display will maintain zero until a change of 0.5 divisions per second has
been exceeded.
1d
= the display will maintain zero until a change of 1 divisions per second has been
exceeded.
3d
= the display will maintain zero until a change of 3 divisions per second has been
exceeded.
Stable Range
[
Stable
]
Set the amount the reading can vary while the stability symbol remains on.
0.5d
= 0.5 balance division
1d
= 1 balance division
2d
= 2 balance division
5d
= 5 balance division
Reset:
no
, yes
Filter:
Low,
Med
, High
Auto Zero Tracking: off,
0.5d
, 1d, 3d
Stable:
0.5d,
1d
, 2d, 5d
Backlight:
off, on,
auto
Auto Tare:
off
, on, on-acc
Auto Off:
off
, 1, 5, 10
End Setup:
Exit menu
Summary of Contents for SJX323
Page 2: ...EN 2 SCOUTTM SJX SERIES BALANCES...
Page 24: ...ES 2 Balanzas de la gama Scout TM SJX Series...
Page 49: ...Balanzas de la gama Scout TM SJX Series ES 25...
Page 74: ...DE 2 SCOUTTM WAAGEN SERIEN SJX...
Page 98: ...IT 2 Bilance Serie Scout TM SJX...
Page 123: ...Bilance Serie Scout TM SJX IT 25...