5. MAINTENANCE
5.1 Cleaning
WARNING:
Electric Shock Hazard. Disconnect the equipment from the power
supply before cleaning.
The housing may be cleaned with a cloth dampened with a mild detergent if necessary.
Attention:
Do not use solvents, chemicals, alcohol, ammonia or abrasives to
clean the housing or control panel.
5.2 Troubleshooting
The following table lists common problems and possible causes and remedies.
If the problem persists, contact OHAUS or your authorized dealer.
TABLE 5-1
Symptom
Possible Cause
Cannot turn on
No power to balance
Poor accuracy
Improper calibration
Unstable environment
Cannot calibrate
Security switch in locked position
Err 8.1
Weight exceeds power on zero range.
Err 8.2
Weight below power on zero range.
Err 8.3
Over load (weight exceeds rated capacity)
Err 8.4
Under load (pan removed)
Err 8.5
Tare out of range
Err 8.6
Displayed value exceeds 999999 (possible in Totalization mode)
Err 8.7
Unknown internal weight position
ref.err
Parts counting or percentage error – sample weight <1d. Balance
shows error then exits parts counting or goes to [
Clr.APU
].
Lo.ref
Percent Reference weight or APW is too low for accurate results
CAL E
Fail to do calibration.
USB.Err
Cannot find menu or app file in U-disk.
Summary of Contents for SJX323
Page 2: ...EN 2 SCOUTTM SJX SERIES BALANCES...
Page 24: ...ES 2 Balanzas de la gama Scout TM SJX Series...
Page 49: ...Balanzas de la gama Scout TM SJX Series ES 25...
Page 74: ...DE 2 SCOUTTM WAAGEN SERIEN SJX...
Page 98: ...IT 2 Bilance Serie Scout TM SJX...
Page 123: ...Bilance Serie Scout TM SJX IT 25...