Ranger™ 2000
Series EN-11
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
]
(Ready) will then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first
weight larger than 5d will be displayed and held.
Semi-automatic operation
(DYNAMIC is set to SEMI in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready)
will then be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second
increments.
3. After completed countdown, the average weight will be displayed. The tilde
symbol will blink indicating that the current weight is being held.
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
]
(Ready) will then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first
weight larger than 5d will be displayed and held.
Automatic operation
(DYNAMIC is set to AUTOMATIC in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready)
will then be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second
increments.
3. After completed countdown, the average weight will be displayed. The tilde
symbol will blink indicating that the current weight is being held.
4. Remove the load and the display will reset after 10 seconds, [
rEAdY
] (Ready)
will then be displayed.
Notes
: If Set0 has been selected, the countdown timer is not displayed. The
first weight larger than 5d will be displayed and held.
Alternatively, the display may be manually reset by pressing the
Function
key.