Ranger 7000
®
Series Service Manual
86
Ohaus Corporation www.ohaus.com
Figure B-7. Replace Loadcell Tab.
Figure B-8. Function selection Replace Loadcell.
15. Configure the Com Port Configuration (B.2) and click 'Next'.
16. Key in the scale serial number, Model number, Temperature Reference in Celsius and select
your region (AP, EU/GB or US).
17. Enter the IDNR number (xxx.xx.xx.xxxx) from the label on the back of the scale or from the
packet label of the new load cell.
18. Click the Start button.