EN-124
EXPLORER BALANCES
5.12.6 Data Maintenance
To check Software Upgrade, Balance Info and System Log.
All the changes made on balance setting will keep records in system log file. e.g. every
printed data, date/time change, balance setting changes, perform calibration action, user
log in/log out, user account create/edit/delete and etc.
These system log files can be reviewed and exported as a PDF to a USB flash drive.
When exporting the system log files, the two touchless sensor's LEDs will blink.
Maximum 5000 records can be stored in balance memories. When the memories are full,
a message will pop up for exporting the log files to a USB flash diver. Then the pop-up
message will show the log file has deleted.
In addition, calibration logs can be viewed separately
in System Log. To do this:
Activate
System Log
selection in the
Date
Maintenance
menu.
The
System Log Review
menu appears. Touch it.
Touch
View Calibration
to view calibration logs.
5.12.7 Service Mode
Used to enter the Ohaus Service menu (Service personnel only). A password is required
to access this menu;