4-6
CHAPTER 4 TESTING
TABLE 4-5. AV MODELS TEST MASS VALUES
Balance Capacity
Mass 1
Mass 2
AV53
220
230
AV212
450
470
AV412
1300
1400
AV812
1300
1400
AV2101
13kg
14kg
AV4101
13kg
14kg
AV8101
13kg
14kg
4.2.4 Load Cell Adjustments (Strain Gauge Models Only) (Cont.)
Up Stop Adjustment
Checking and adjusting the up stop.
CAUTION
DO NOT PULL UP ON THE LOAD CELL AS DAMAGE
MAY OCCUR. USE FEELER GAUGES TO CHECK
CLEARANCES.
1. Using a feeler gauge, check the clearance of the up stops. See Tables 4-6 or 4-8.
2. If the clearance is excessive, adjust the nuts using the data in Tables 4-7 or 4-8. See
Figure 4-3. The screw with the nuts should be held fast while adjusting the nuts. Locktite
the nuts when completed. Do not pull up on the Load Cell.
TABLE 4-6. AS MODELS TEST MASS VALUES
Balance Capacity
Mass 1
Mass 2
AS153
220
230
AS312
450
470
AS612
1300
1400
AS811
1300
1400
AS1502
2.2kg
2.5kg
AS3101
13kg
14kg
AS6100
13kg
14kg
AS8100
13kg
14kg