![OHAUS AV114C Service Manual Download Page 49](http://html1.mh-extra.com/html/ohaus/av114c/av114c_service-manual_726686049.webp)
3-25
CHAPTER 3 REPAIR PROCEDURES
3.2.11
Unmounting the Load Cell from the Base Plate
This procedure is necessary before proceeding with other repairs to the Load Cell.
Base Plate Removal
1. Refer to Figure 3-36 and remove the Weigh Below Hook and screw.
Weigh Below Hook and Screw
Mounting Screws
Figure 3-36. Base Plate Removal.
2. Hold the MFR Load Cell and remove the 2 Load Cell Mounting Screws. Figure 3-37
illustrates the components.
Figure 3-37. Load Cell Mounting Components.
MFR Load Cell
Base Plate
Weigh Below Hook and Screw
Mounting Screws
Note: At this point, other repairs may be made to the Load Cell.
Mating Surfaces