Problem
Sign
Cause
Solution
Dusting of Sieve Beds
White powder visible
in the machine or very
high air pressure levels
Contaminated sieve
Moisture in air
Repack the sieve beds.
Clean valves and mufflers.
Make sure dry air is being
fed to the machine.
Valves Sticking
Air pressure levels
too high
Dusting of sieve or
machine filled with dirt
and dust due to air
filters not being
replaced
Remove valve block from
machine and clean valves
and spools completely.
Pressure Switch not
functioning
Machine not turning
ON/OFF
at target
pressures
Faulty switch
Remove switch and return
for replacement.
Warning Signs:
Low Oxygen Pressure
This may be a result of a leak in the system.
Use a leak testing solution to locate and repair any
air leaks.
The machine has run for 30 minutes and purity has
not yet been reached
This may be a result of a leak in the system.
Use a leak testing solution to locate and repair any
air leaks.
Oxygen purity has fallen below acceptable limits
This may be an indication of a leak within the system.
Use a leak testing solution to locate and repair any leaks.