32
Example Softcontrol Data/F
SPS:
byte
value
(decimal)
meaning
comment
0
117
Start identifier
1
48
Start identifier
2
0
Status word Bits 8..15
n/a
3
32
Status word Bits 0..7
Bit 5 set
device (sensor) alarm
4
3
Unit of measurement
3 = ppm
5
1
Language
1 = English
6
0
Measurement counter (Hi)
28
28 soft water measurements
have already been carried out
7
28
Measurement counter (Lo)
8
0
Raw water hardness
3*256 + 232 = 1000
RW
= 180 ppm
9
0
Raw water hardness
10
3
Raw water hardness
11
232
Raw water hardness
12
0
Soft water hardness
13
SW = 2.34 ppm
13
0
Soft water hardness
14
0
Soft water hardness
15
13
Soft water hardness
16
0
Soft water limit value
20
3.60 ppm
17
0
Soft water limit value
18
0
Soft water limit value
19
20
Soft water limit value
20
0
Sensor quality
120
S = 12
21
120
Sensor quality
22
1
Measurement pause
1*256 + 44 = 300
Measurement pause = 5 minutes
23
44
Measurement pause
24
2
Alarm delay
2*256 + 88 = 600
Alarm delay = 10 minutes
25
88
Alarm delay
26
0
n/a
27
0
n/a
28
0
n/a
29
0
n/a
30
54
6
Sensor number 6370214003
31
51
3
32
55
7
33
48
0
34
50
2
35
49
1
36
52
4
37
48
0
38
48
0
39
51
3