12
Software
Setup
Select “Setup” from the “Options” menu.
Minimum Save Deltas:
These fields determine how often data is saved to
the file. A data point will be recorded when any of the values changes by the
amount specified.
Time:
To set the save period to one minute, set the value in this field
to “01:00”.
Temp:
If the value in this field is set to 1, a data point will be recorded
whenever the temperature changes by more than 1°. Units are
determined by the “Temp Unit” field below.
Distance:
If the value in this field is set to “0.0010”, a data point will be
recorded whenever the corrected distance changes by more
than 0.001. Units are determined by the “Distance Unit” field
below.
Swell:
To record a data point whenever the swell % change by .01%,
set the value in this field to .01.
Number of Available Cells:
By default, the Swell Meter can operate
four cells at a time. If your Swell Meter has an extra control card installed
(optional), change this value to eight.
Temp Unit / Distance Unit:
Choose the temperature unit (°F or °C) and
distance unit (in or mm) you wish to use.
Archive Path:
Select a folder to store test data.
Print to Printer:
When a test prints, the software automatically creates an
image file of the chart. If the Print to Printer open is selected, it will also print
to the default printer.