OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
11
Software
Calibration
1.
From the “Edit” menu, choose “Calibration”. The following fields make up
the calibration screen:
Variable:
Selects which device is being calibrated.
Value Low:
The low value in the calibration
Value Hi:
The high value in the calibration
Sig. Lo:
The signal value corresponding to
Value Low
.
Sig. Hi:
The signal value corresponding to
Value Hi
.
Raw Value:
The value currently being sent by the equipment.
Low (Button):
Copies the
Raw Value
into the
Sig. Lo
. field.
High (Button):
Copies the
Raw Value
into the
Sig. Hi
. field.
Store Cal:
Stores the calibration values in memory.
2. Choose the device to be calibrated from the “Variable” drop-down list.
a.
Sample Temp
and
Cell Temp
i. Before calibrating either temperature, set the temperature units to
°C. Refer to page 2 for instructions.
ii. Insert the thermocouple into a calibrating block. Make sure the
thermocouple is plugged into the appropriate port.
iii. Set the calibrating block to 0°C. Refer to the documentation
provided with the calibrating block for instructions.
iv. Enter the reading from the calibrating block into the
Value Low
field. Once the
Raw Value
has stabilized, click the “Low” button.
v. Set the calibrating block to 200°C.
vi. Enter the reading from the calibrating block into the
Value Hi
field.
Once the
Raw Value
has stabilized, click the “High” button.
vii. Click the “Store Cal” button to save the calibration data.