Lubricity Test
Refer to page 4 for instructions on using the control panel.
1. Clean the lubricity test ring (#111-02 with a flat outer surface) and the
lubricity test block (#111-08 with a concave groove on one side) with
Acetone and rinse them thoroughly with deionized water. All parts of
the machine in the sample area must be clean before starting a test.
Do not touch the metal contact areas with bare hands.
Removing the ring from the main shaft can change the alignment of the
contact area with the block. We recommend removing the ring only to
replace it or to switch between lubricity and extreme pressure tests.
2. Place the lubricity test ring squarely onto the tapered portion of the
main shaft. Using a
15
⁄
16
" wrench, secure the test ring retainer nut.
Make sure the ring seats squarely on the taper of the shaft.
3. Turn on the power and let the machine run for approximately 15 min-
utes.
4. Place the lubricity test block in the block holder with the concave side
facing out and align it with the test ring.
5. Set the motor speed to 60 RPM.
6. After the unit has been running for 15 minutes, zero the torque reading
(refer to page 4 for instructions). Run the unit approximately 5 more
minutes and adjust again if required.
7. Fill the stainless steel sample cup with deionized water (260 - 280 mL)
and place it on the lowered cup stand. Raise the cup stand until the
test ring, test block, and block holder are fully submerged. Tighten the
thumbscrew to secure the cup stand.
OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
5
Sample Cup
Retainer Nut
Test Ring
Main Shaft
Thumb Screw
Cup Stand
Locking Nut
Torque
Bushing
Important
!
Note