Page 35 of 35
Odom Hydrographic Systems, Inc.
May 28, 2008
The parameter IDs that you use to send settings are the same as when you request settings. For example, to set
the chartspeed you would use a parameter ID '15'.
Querying Parameter settings
To query a setting use the following 6 character sequence: <CTRL+P><Request ID><space><Parameter ID>
The Request ID is a command that the Echotrac recognizes as the command for a parameter value request.The
Echotrac CV will then reply with “#CV3,P,”<UNITS><PING NUMBER><PARAMETER ID><VALUE>.
The Units is a single character F or M for Feet or Meters. The ping number is 8 characters, the Parameter ID is 2
characters and the value is 8 characters. Example:
<CTRL+P>86 00
#CV3,P,F000000010000000168
Note:
The Serial Parameter Request feature is supported in the following firmware versions only (See table
below). Also, as of version 3.06 the ID used to request a parameter value changed from 86 to 187.
Firmware
Request ID
3.01 – 3.05
86
3.06 –
187
See chapter 7 Overview parameters and settings for a table listing all the parameters and their unique identifier.