10 Troubleshooting
Instruction Manual BIOM® 5 (G/55400/1218/en)
23 / 36
10 Troubleshooting
Attention
If a fault occurs which you are unable to correct by following the
instructions below, label the device as “out of order” and contact our
service department. (Address:
).
Never operate a damaged BIOM® 5.
Before returning to OCULUS: Condition the BIOM® 5 and its
components in accordance with the conditioning instructions.
Troubleshooting guide – BIOM® 5
Fault
Possible cause
Remedy
The safety rod of the BIOM® 5 is
stuck
BIOM® 5 must not be used in this
state!
Deposits on the BIOM® 5 due
to inadequate sterilisation
Foreign body in safety rod
extension channel
Safety rod is bent
Carefully mechanical
cleaning and change to a
better demineralised water,
clean the BIOM® 5
ultrasonically
Careful mechanical cleaning
and removal of the foreign
body
Clean the BIOM® 5 in the
ultrasound bath
Treat the rod with a suitable,
silicone oil-free lubricant
prior to the next sterilisation
Send the BIOM® 5 to OCULUS
Service
Loose adaptor
Dovetail mount wobbles
The locking screws are loose
Screws are loose
Tighten the locking screws by
hand
Tighten the screws with a
suitable screwdriver
Image is cropped or out-of-center
The SDI® or BIOM®
connecting spacer , other
components or the BIOM® 5
adapter are incorrectly
mounted at an angle
The magnifying lens clip is
bent or mechanically
damaged
Correct the assembly
Send the loupe to
OCULUS Service for
readjustment