10 Troubleshooting
Instruction Manual BIOM® 5 (G/55400/1218/en)
25 / 36
No function whatsoever when the
combination control unit is actua-
ted
The combination control unit
is not connected to the
SDI® 4c or BIOM® connecting
spacer
The SDI® 4c or BIOM®
connecting spacer is not
connected to the 6V-15V
power supply
Power failure or power socket
is not active
These are not active when the
sockets at the microscope
stand are in use
Establish the connection to
the SDI® 4c or BIOM®
connecting spacer
Establish the connection to
the 6V-15V power supply
Inform the in-house
electrician
Use the 6V-15V plug
transformer
Use the mechanical adjusting
element or adjusting wheel
Activate the sockets in
accordance with the
instructions for the stand
Contact the microscope
manufacturer for assistance
Malfunction when using the
combination control unit
5-pole plug has been forcibly
plugged in the wrong way
round
Plug it in the right way round
(pay attention to the lug and
slot of the polarity reversal
protection)
Motorized focussing not possible with
the
BIOM® 5
c/cl when using the
combination control unit
BIOM® 5c/cl connector not
plugged into the SDI® 4c or
BIOM® connecting spacer
properly
Defective drive belt
Drive belt missing
Connecting cable damaged
Defective drive module
Plug in the connector
correctly
Install a new, sterile drive
belt, or focus manually using
the focussing knob at the
BIOM® 5c/cl
Install a sterile drive belt
Exchange the drive module
Exchange
Fault
Possible cause
Remedy