9 Use of the BIOM® 5
Instruction Manual BIOM® 5 (G/55400/1218/en)
21 / 36
Attention
Incorrect detaching is hazardous to the patient
Never detach the BIOM® 5 over the patient.
To dismount the BIOM® 5, swing it to its standby position.
BIOM® 5m/ml
Pull the BIOM® 5 off the adaptor plate.
Loosen the knurled screw,
.
Pull the adaptor plate off the dovetail mount.
Bring the BIOM® 5 and its components to the conditioning station
immediately.
BIOM
®
5c/cl
Unplug the connector of the BIOM® 5c/cl from the socket at the
SDI® 4c or BIOM® connecting space housing. To unlock the plug,
take hold of the outer plug sleeve.
Fig. 9-10: Pull off the adaptor plate
Fig. 9-11: BIOM® 5c/cl: Disconnect the plug-in connectors