9 Use of the BIOM® 5
20 / 36
Instruction Manual BIOM® 5 (G/55400/1218/en)
Special Case „Air-filled eye“
During the fluid-air-exchange: Turn the focussing knob of the
BIOM® 5 until the front lens reaches the highest position.
Use the microscope’s fine focus to enlarge the image section.
For the BIOM
®
5c/cl only (focussing with electric motor):
Focussing of the BIOM® 5c/cl is done by the surgeon by means of the
combination foot switch while observing through the microscope.
Only use the BIOM® 5c/cl’s motorized focussing function when the
front loupe is far enough away from the patient’s eye.
The surgeon may only use the motorized focussing function when
the distance between the ophthalmoscopy loupe and the eye is
simultaneously
monitored.
It must be ensured that the surgeon can stop the motorized
focussing function at any time.
Attention
Risk of eye damage if the working distance between the BIOM® 5 and the
patient changes
When the BIOM® 5 is swung into the working position (into the beam pa-
th), the following instructions must be followed:
Do
not
use the coarse adjustment knob at the microscope stand to
adjust the height.
Do
not
adjust the height of the stand arm, either by motor or
manually, over the surgical area.
Do
not
change the patient’s position by adjusting the height of the
operating table.
9.8
During the Surgery
Make sure that the cornea is sufficiently moistened with a suitable
solution.
9.9
After the Surgery
After the surgery, you must remove the BIOM® 5 from the microscope.
You must bring the BIOM® 5 and its components to the conditioning
station immediately after the surgery.