MNL-1007 REV A
22
o
Stop
adjusting
the distance
at the point
where the
value of the
peak to peak
centerburst
in the
interferogram and/or the PSD value are maximum and lock the probe in position. This is the optimum optical
distance between the probe and reference/standard.
o
Optional: Capture the plots and save them for future reference.
•
Click
Stop
.
3.
Set measurement parameters:
•
Go to the Spectrum tab.
•
Scan Time:
Set the scan time that best fits your requirements. The higher the Scan Time, the higher the number of
acquisitions averaged and the better the signal to noise ratio.
•
Resolution: Select either 8 or 16 nm optical resolution (FWHM).