Installing a Needle

To place the gutta percha where desired in the root canal, choose an applicator needle diameter best suited (20, 23 or 
25-gauge). With the Obtura III MAX off and cool, remove the needle nut using the Multi-Tool; the needle nut should 
remain on the Multi-Tool when unscrewed. Place the appropriate needle in the rear of the needle nut, so that the length of 
the needle protrudes through and forward (diagram G). Re-attach the needle nut on to the heater thread using the Multi-
Tool, making sure not to cross-thread (diagram H). Tighten until snug (do not over tighten, but do not leave too loose).

Bending the Needle

In most situations, a straight applicator needle will not provide access to the root canal. Using the Multi-Tool, gently 
bend the needle to provide a shape conducive to the procedure (see diagram I). Be careful not to bend too sharply as 
kinks will result and gutta percha flow will be blocked. For access to difficult to reach root canals, simply loosen the 
needle nut, rotate the needle to desired position, and then re-tighten.

Loading Gutta Percha

WARNING: USING MORE THAN ONE STICK OF GUTTA PERCHA AT A TIME MAY DAMAGE THE SEAL. 
DO NOT USE MORE THAN ONE PIECE AT A TIME.

Power on the Obtura III MAX and wait for the heater to come up to your desired temperature. Press the handpiece release 
button on the top of the Obtura III MAX handpiece and withdraw the plunger shaft. Insert a gutta percha stick into the 
chamber via the opening just forward of the release button (see diagram J). Re-insert the plunger shaft until the plunger seal 
engages the gutta percha pellet. Some resistance will be felt as the gutta percha is pushed forward to the heating chamber. 

Expressing Gutta Percha 

Specifics to technique will not be addressed here, but a few reminders and hints follow:

         

 When expressing gutta percha, do not “

withdraw

” the Obtura III MAX, but let the gutta percha fill “

push

” the

----------

needle out of the canal. Pulling the needle out may cause voids and require condensation or re-work. 

         

 Take care when bending needles, as a kink will make it all but impossible to express material. Also, repeated

----------

bending will cause needle weakening and subsequent breakage. 

         

 Wait for the gutta percha to come up to full operating temperature before attempting to express; material that

----------

is too cool will require much more pressure or may be impossible to pass through a small diameter applicator needle. 

         

 Before beginning a procedure, express about 3 cm of gutta percha so that the up-to-temperature material will be

----------

at the end of the needle; the warm material will fill space more completely.

INSTALLING A NEEDLE       

5

G

I

J

H

OPERATOR’S

MANUAL

HEATED

 GUTTA PERCHA 

SYSTEM

Factory Repair/Contact Us

Factory Service is available at Obtura Spartan. Please call the number below for technical support and 
problem diagnosis.

Obtura Spartan 
(800) 344-1321
www.obtura.com

It is necessary to acquire an RMA number before returning your unit. Failure to do this may result in longer repair

 

times and additional charges. 

If you have spoken to the repair department and it is determined that the unit must be returned to the factory, 
please be sure to pack carefully and appropriately.  Use the original box and packaging (if possible), or a 
minimum of 8 cm (3 in) impact-absorbing insulation in your new box. Obtura Spartan will not be responsible for 
damage due to improper packing or shipping. Insurance is optional and highly recommended.  

Note:

 Unauthorized repair will void all warranties. Authorized repair procedures must follow all safety 

precautions. Be sure that the power supply connector is disconnected from the power supply port before 
authorized repairs, and that safety glasses are worn at all times during the authorized repair procedures.

Replacement Parts

PN

PN

Description

Description

823-800

823-810

823-711

823-616

823-618

823-812

823-620

823-635

823-815

822-602

822-604

823-813

822-613

825-103

823-814

823-623

822-609

Obutra III MAX Complete System

Obtura III MAX Handpiece

Obtura Power Supply

Obtura Needle Nuts - Pack of 2

Obtura Plunger Seal Assembly

Obtura III MAX Plunger Shaft with Plunger Seal

Obtura Applicator Needles - 20 Gauge

Obtura Applicator Needles - 25 Gauge

Obtura III MAX Thermal Protectors - Pack of 4

Obtura Gutta Percha - Box of 100 Pieces

Obtura Flow 150 Gutta Percha - Box of 100 Pieces

Obtura III MAX Multi-Tool

Obtura Cleaning Brushes - Pack of 2

Obtura Heater Reamer Tool

Obtura III MAX Handpiece Bushing

Obtura Applicator Needles - 23 Gauge

Obtura Cleaning Kit - Bottle of Cleaning 

Solution with 2 Brushes

12

   

FACTORY REPAIR/CONTACT US

Summary of Contents for III MAX

Page 1: ...ATOR S MANUAL Welcome to the Your essential guide to our clinically advanced Heated Gutta Percha System 2260 Wendt Street Algonquin IL 60102 800 344 1321 www obtura com 2011 Obtura Spartan 57412G 10 1...

Page 2: ...supplied by Obtura Spartan only This equipment contains no electrical parts serviceable by the user Environmental Conditions Symbol SN PN M Symbol Description Description Ambient Temperature 10 C to...

Page 3: ...ha System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Obtura III Max Heated Gutta Percha System can help prev...

Page 4: ...equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the...

Page 5: ...ies Use of other cables or accessories has the potential to damage the device or affect compliance to EMC standards 9 Cables and Accessories Handpiece Cord Max length 8 IEC 60601 1 Power Cord Max leng...

Page 6: ...and difficult to move the plunger due to a buildup of old hardened gutta percha Overview Congratulations on your purchase of the Obtura III MAX Please read the entire manual prior to using this unit H...

Page 7: ...cause irritation of the respiratory tract y nausea dizziness or headache Product has a low order of acute oral toxicity but ingestion of large quantities may cause nausea vomiting and gastrointestinal...

Page 8: ...ge from the previously set temperature to the new set point temperature Note preset memories are saved even during power failure The Obtura III MAX requires approximately 120 seconds to heat from ambi...

Page 9: ...t for the gutta percha to come up to full operating temperature before attempting to express material that is too cool will require much more pressure or may be impossible to pass through a small diam...

Page 10: ...ol unit case via a standard power supply connector Frequently Asked Questions Q How often should the unit be cleaned A The Obtura III MAX should be cleaned at the end of each day or procedural session...

Page 11: ...pring loaded pawl moves the shaft forward The heater assembly is directly connected to the handpiece shell providing a straight through path for gutta percha warming it in the process 10 ERROR CODES M...

Page 12: ...that bushing rotation ceases To install the handpiece bushing slide it in the opening at the rear of the handpiece diagram K and turn clockwise until resistance is felt The handpiece bushing should b...

Reviews: