Error Codes

Handpiece Attachment/Detachment

To Attach: Hold the handpiece cord connector at the rear of the main control unit, facing the triangle indicator at the 
12 O’ Clock position (see diagram N). Plug in to the handpiece cord port when proper alignment is reached.

To Detach: Squeeze the serrated section on the handpiece cord connector (by the triangle) and gently pull back (see 
diagram O). Connector does not turn.

Error Code

Fault

Remedy

S-1

S-2

H-1

H-2

H-3

B-1

P-1

Handpiece disconnected (communication failure)

Cannot reach Temp. in 255 sec

Heater Wire open

Heater Wire short

Thermocouple error

Internal control base failure

Power Supply fault

Heater fault – Contact factory for repair

Replace handpiece or contact factory for repair guidance

Replace handpiece or contact factory for repair guidance

Replace handpiece or contact factory for repair guidance

Not user-serviceable – Contact factory for repair

Replace power supply or contact factory for repair guidance

1) Verify fitment of handpiece connector

2) Remove and replace

3) Check cable for breaks or damage

4) Check connector for broken or bent pins

5) Contact factory for handpiece repair/replacement

N

O

OPERATOR’S

MANUAL

HEATED

 GUTTA PERCHA 

SYSTEM

Theory of Operation 

The heating system of the Obtura III MAX is a closed-loop design to ensure consistent temperatures and uniform gutta 
percha flow. This means that the temperature of the heater is read several hundred times per second and corrected (as 
necessary) to insure accurate and consistent heating. 

In the handpiece, the temperature of the heating chamber is detected by a thermocouple integrated into the heating 
element and encapsulated with epoxy. The output from the thermocouple is converted by an analogue-to-digital 
converter, and into a digital stream for use by the main control unit.

The handpiece is composed of a two-piece heat-resistant composite body that encapsulates the trigger assembly, heater 
system and plunger.

Gutta percha is placed into the top of the handpiece shell and expressed via the plunger shaft assembly. Pressing the 
handpiece trigger, which pushes the shaft via a spring-loaded pawl, moves the shaft forward. The heater assembly is 
directly connected to the handpiece shell, providing a straight through path for gutta percha, warming it in the process.

10

   

ERROR CODES 

Maintenance

For best results, the Obtura III MAX handpiece should be cleaned daily. This prevents build-up of excess gutta percha 
within the heating chamber. 

Regular Maintenance (Day to Day Cleaning)*

Tools you will need:

 

 

Multi-Tool

 Cleanin

  

g brush and  Cleaning solution

 Disposab

  

le cloth (na pkin, gauze, etc.)

1. Prior to cleaning, ensure that your Obtura III MAX unit is turned off, fully cooled and unplugged.

2. Unlatch and remove the thermal protector using the Multi-Tool.

3. A) Use the Multi-Tool to unscrew the needle nut from the heater thread.

B) Set aside the needle nut. Dispose of the needle (for optimum performance, Obtura applicator needles

------

should be used only once).

4. Extrude all remaining gutta percha by squeezing the handpiece trigger. The gutta percha will come out in solid pellet form.

5. Push and hold the release button, removing the plunger shaft from the handpiece.

6. A) Inspect the plunger seal for gutta percha.

B) If gutta percha is present, dip the cleaning brush into the cleaning solution, fully coating the brush.
C) Clean the plunger seal with the cleaning brush in a back and forth motion until all visible gutta percha is cleared.

7. A) Inspect the inside of the handpiece shaft for any visible gutta percha.

B) If gutta percha is present, re-dip the cleaning brush into the cleaning solution, fully coating the brush.

C) Place the cleaning brush inside the handpiece shaft and clean in a back and forth motion until all visible

------

gutta percha is cleared.

8. A) Reassemble all components in the reverse order from which they were detached.

B) Place the handpiece back on its cradle.

WARNING: AT NO POINT SHOULD ANY PART OF THE OBTURA III MAX HANDPIECE BE SUBMERGED 
INTO THE CLEANING SOLUTION.

NOTE:

 A) For optimal cleaning, use a Birex®-type solution to aid in disinfecting and deodorizing your Obtura III MAX unit.

B) Never spray the disinfectant solution directly on to the unit; always spray into a disposable cloth, then wipe down.

*The above regular maintenance cleaning instructions are also displayed on the provided laminated cleaning
instruction card, which contains visual diagrams for your convenience.

MAINTENANCE       

7

Summary of Contents for III MAX

Page 1: ...ATOR S MANUAL Welcome to the Your essential guide to our clinically advanced Heated Gutta Percha System 2260 Wendt Street Algonquin IL 60102 800 344 1321 www obtura com 2011 Obtura Spartan 57412G 10 1...

Page 2: ...supplied by Obtura Spartan only This equipment contains no electrical parts serviceable by the user Environmental Conditions Symbol SN PN M Symbol Description Description Ambient Temperature 10 C to...

Page 3: ...ha System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Obtura III Max Heated Gutta Percha System can help prev...

Page 4: ...equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the...

Page 5: ...ies Use of other cables or accessories has the potential to damage the device or affect compliance to EMC standards 9 Cables and Accessories Handpiece Cord Max length 8 IEC 60601 1 Power Cord Max leng...

Page 6: ...and difficult to move the plunger due to a buildup of old hardened gutta percha Overview Congratulations on your purchase of the Obtura III MAX Please read the entire manual prior to using this unit H...

Page 7: ...cause irritation of the respiratory tract y nausea dizziness or headache Product has a low order of acute oral toxicity but ingestion of large quantities may cause nausea vomiting and gastrointestinal...

Page 8: ...ge from the previously set temperature to the new set point temperature Note preset memories are saved even during power failure The Obtura III MAX requires approximately 120 seconds to heat from ambi...

Page 9: ...t for the gutta percha to come up to full operating temperature before attempting to express material that is too cool will require much more pressure or may be impossible to pass through a small diam...

Page 10: ...ol unit case via a standard power supply connector Frequently Asked Questions Q How often should the unit be cleaned A The Obtura III MAX should be cleaned at the end of each day or procedural session...

Page 11: ...pring loaded pawl moves the shaft forward The heater assembly is directly connected to the handpiece shell providing a straight through path for gutta percha warming it in the process 10 ERROR CODES M...

Page 12: ...that bushing rotation ceases To install the handpiece bushing slide it in the opening at the rear of the handpiece diagram K and turn clockwise until resistance is felt The handpiece bushing should b...

Reviews: