Obtura III MAX Operator'S Manual Download Page 6

Overview

Congratulations on your purchase of the Obtura III MAX! Please read the entire manual prior to 
using this unit. Here are some useful tips that will aid your usage of the Obtura III MAX:

When you are ready to practice with your Obtura III MAX, you should become familiar with its components. 
After attaching the needle and thermal protector, turn the Obtura III MAX on and set it to operating temperature. 
It takes the unit approximately 120 seconds to reach the highest operating temperature of 200°C. The high 
temperature setting will improve gutta percha flow, especially when used in the cold practice block (sold 
separately). Begin by filling the plastic practice blocks with gutta percha per the instructions included. Use 
condensers to condense the material void free. We recommend the Obtura Spartan S-Kondenser (available in 3 sizes).

 

 

The soft silver applicator needles are designed to conduct heat to insure 

 even at the needle tip; however,

as the needle tip cools between uses, you should express 3 cm of gutta percha to pre-heat the tip just prior to
use. The needles are bendable to fit the tooth, however, they may break if they are repeatedly bent or if they are
bent at the needle attachment point. Always use the Multi-Tool to bend the needle. Do not kink the needle, as
any restriction will impede flow. After a few practice sessions, proper needle bending will be easily accomplished.
Obtura applicator needles are designed for one use only.

 

 

Do not loosen or remove the blue plastic Heater Nut. The Heater Nut is fastened to the gun with a set-screw; 

removal of the heater nut will cause damage. The Heater Nut must be kept tightly fastened in order to secure the
heater to the handpiece when expressing gutta percha.

 

 

The stand for the handpiece may be positioned by itself or it may be attached to the bottom of the control

unit. The stand can be attached to either side of the control unit with four phillips-head screws that are fastened
to the bottom of the stand (see page 3 for further details).

 

 

The cord to the handpiece can extend from the front or rear of the unit. If a front exit is desired, gently press the

handpiece cable into the tracks on the bottom of the unit, to exit either on the left-front or the right-front (page 3).

 

 

Cleaning the Obtura III MAX is simple but very important; just follow the directions in this manual (page 6)

or on the provided laminated cleaning instruction card. If your Obtura III MAX is cleaned on a regular basis it
will provide years of reliable service. If you do not clean your unit, it may become clogged and difficult to move
the plunger due to a buildup of old, hardened gutta percha.

Overview

Congratulations on your purchase of the Obtura III MAX! Please read the entire manual prior to using this 
unit. Here are some useful tips that will aid your usage of the Obtura III MAX:

Thank you for purchasing the Obtura III MAX. We believe it is a beneficial instrument that will enhance 
your practice. We are continually striving to improve our products and services. If you have any questions or 
suggestions, we would appreciate hearing from you. We are also interested in seeing any X-rays you may have 
of interesting cases. Feel free to contact us at: 1-800-344-1321 (or 1-877-485-3556 in Canada); we welcome the 
opportunity to speak with you. Also, visit our website at: www.obtura.com.

System Description

The Obtura III MAX is intended for use by professionally-qualified dentists, endodontists and dental clinicians to heat 
gutta percha and to place it into the previously prepared root canals of human teeth, in order to provide quick and 
complete obturation of the root canal. 

There is a minimum level of understanding and background assumed in the descriptions and instructions found 
in this manual; if at any time you feel overwhelmed or unsure of how to properly operate your Obtura III MAX 
unit, immediately cease use and call Obtura Spartan for assistance. We value safety first and foremost.

The Obtura III MAX features an interchangeable plug-in handpiece. Any Obtura III MAX handpiece will work with 
any Obtura III MAX base unit. However, the Obtura III MAX handpiece cannot be retrofit to the Obtura III or any 

2

   

OVERVIEW

OPERATOR’S

MANUAL

HEATED

 GUTTA PERCHA 

SYSTEM

GUIDANCE AND MANUFACTURER'S DECLARATION

    

1

5

IEC 60601-1-2:2007 (Ed 3.0) 

Guidance and Manufacturer’s Declaration – Electromagnetic Emissions 

The Obtura III Max Heated Gutta Percha System is intended for use in the electromagnetic environment specified 

below. The customer or the user of the Obtura III Max Heated Gutta Percha System should assure that it is used in 

such an environment. 

Emission test 

Compliance 

Electromagnetic environment 

– guidance 

RF emission 
CISPR 11 

Group 1 

The Obtura III Max Heated Gutta Percha System uses RF 
energy only for its internal function. Therefore, its RF 
emissions are very low and are not likely to cause any 
interference in nearby electronic equipment. 

RF emissions 
CISPR 11 

Class B 

The Obtura III Max Heated Gutta Percha System is 
suitable for use in all establishments, including domestic 
establishments and those directly connected to the public 
low-voltage power supply network that supplies buildings 
used for domestic purposes. 

Harmonic emissions 
IEC 61000-3-2 

IEC 61000-3-2 class A 

Voltage fluctuations/flicker 
emissions 
IEC 61000-3-3 

Complies 

EMC Table

Summary of Contents for III MAX

Page 1: ...ATOR S MANUAL Welcome to the Your essential guide to our clinically advanced Heated Gutta Percha System 2260 Wendt Street Algonquin IL 60102 800 344 1321 www obtura com 2011 Obtura Spartan 57412G 10 1...

Page 2: ...supplied by Obtura Spartan only This equipment contains no electrical parts serviceable by the user Environmental Conditions Symbol SN PN M Symbol Description Description Ambient Temperature 10 C to...

Page 3: ...ha System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Obtura III Max Heated Gutta Percha System can help prev...

Page 4: ...equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800 MHz to 2 5 GHz where P is the maximum output power rating of the...

Page 5: ...ies Use of other cables or accessories has the potential to damage the device or affect compliance to EMC standards 9 Cables and Accessories Handpiece Cord Max length 8 IEC 60601 1 Power Cord Max leng...

Page 6: ...and difficult to move the plunger due to a buildup of old hardened gutta percha Overview Congratulations on your purchase of the Obtura III MAX Please read the entire manual prior to using this unit H...

Page 7: ...cause irritation of the respiratory tract y nausea dizziness or headache Product has a low order of acute oral toxicity but ingestion of large quantities may cause nausea vomiting and gastrointestinal...

Page 8: ...ge from the previously set temperature to the new set point temperature Note preset memories are saved even during power failure The Obtura III MAX requires approximately 120 seconds to heat from ambi...

Page 9: ...t for the gutta percha to come up to full operating temperature before attempting to express material that is too cool will require much more pressure or may be impossible to pass through a small diam...

Page 10: ...ol unit case via a standard power supply connector Frequently Asked Questions Q How often should the unit be cleaned A The Obtura III MAX should be cleaned at the end of each day or procedural session...

Page 11: ...pring loaded pawl moves the shaft forward The heater assembly is directly connected to the handpiece shell providing a straight through path for gutta percha warming it in the process 10 ERROR CODES M...

Page 12: ...that bushing rotation ceases To install the handpiece bushing slide it in the opening at the rear of the handpiece diagram K and turn clockwise until resistance is felt The handpiece bushing should b...

Reviews: