6.5 Lubrication Schedule
LUBRICANT
TABLE SECTION
MINIMUM FREQUENCY
Grease,
See Table 2-2
Articulation pins
– Leg Section
2 x per year
Articulation pins
– Tabletop Tilting
Articulation pins
– Back Section
Articulation pins
– Seat Section
Central column
– Tabletop up/down
Floor locks
6.6 Schedule for Other Maintenance
ITEMS
MINIMUM FREQUENCY
1. TABLETOP
1.1 Remove pads, and then look for wear and tear on the pads and
on the fastening strips of pads and tabletop.
4× per year
1.2 Examine all side-rail hardware attachments.
1.3 Check and tighten screws on side rails.
1.4 Remove and clean cassette top of each table section.
2. HYDRAULIC SYSTEM
2.1 Clean fluid filters.
1× per year
2.2 Check the cleanness of hydraulic oil. Replace it if necessary.
2.3 Check hydraulic fluid level.
2.4 Inspect floor directly beneath the table and all the hoses, fittings
and components of hydraulic system for leakage.
3. ELECTRICAL CHECKS
3.1 Check for loose circuit board connectors and cable plugs.
1× per year
3.2 Check for damage or fraying cables.
3.3 Inspect electrical cord and Hand Pendant cable.
3.4 Electrically conductive properties of floor locks
4× per year
Summary of Contents for B2AV800 Series
Page 2: ......
Page 18: ...B2AV800 Series Surgical Table IM B2AV800 E01 1 10 This page intentionally left blank...
Page 32: ...B2AV800 Series Surgical Table IM B2AV800 E01 3 8 This page intentionally left blank...
Page 42: ...B2AV800 Series Surgical Table IM B2AV800 E01 4 10 This page intentionally left blank...
Page 50: ...B2AV800 Series Surgical Table IM B2AV800 E01 5 8...
Page 82: ...B2AV800 Series Surgical Table IM B2AV800 E01 8 2 This page intentionally left blank...
Page 84: ...B2AV800 Series Surgical Table IM B2AV800 E01 9 2 This page intentionally left blank...
Page 85: ......