B2AV800 Series Surgical Table
IM-B2AV800-E01
5-13
Display/Code
Symptom
Cause
13
When rotary valve keeps turning but
receives no signal from U1~U4 of
ASM0027
1. Short circuit within the angle
system.
2. Rotary valve motor damaged
3. ASM0023 driver PCB
damaged
4. ASM0027 or its cable
damaged
14
Rotary valve position number not
matching position code
ASM0027 or its cable damaged
15
Rotary valve position code correct but
stops too early, and fails to position
after programmed compensation.
Timing parameter is set too low
16
Rotary valve position code correct but
stops too late, and fails to position
after programmed compensation.
Timing parameter is set too high
17
Not used
18
Not used
19
Not used
20
Not used
21
Batteries are not being charged (i.t.)*
1. Loose battery connector
2. Battery reached life span or
over discharged
3. Broken PCB (ASM0021)
22
Not used
23
Using LOCK or UNLOCK button of
B2AV800L(S) hand pendant on
B2AV800(S)(K) table
1.
Wrong hand pendant
2.
Wrong DIP setting
24
Using Slide button of hand pendant on
Non-slide table
1.
Wrong hand pendant
2.
Wrong DIP setting
25
Using T1000 Series hand pendant on
B2AV800 Series tables.
Wrong hand pendant.
26
Not used
27
Voltage is too high (i.t.)*
Wrong setting
Summary of Contents for B2AV800 Series
Page 2: ......
Page 18: ...B2AV800 Series Surgical Table IM B2AV800 E01 1 10 This page intentionally left blank...
Page 32: ...B2AV800 Series Surgical Table IM B2AV800 E01 3 8 This page intentionally left blank...
Page 42: ...B2AV800 Series Surgical Table IM B2AV800 E01 4 10 This page intentionally left blank...
Page 50: ...B2AV800 Series Surgical Table IM B2AV800 E01 5 8...
Page 82: ...B2AV800 Series Surgical Table IM B2AV800 E01 8 2 This page intentionally left blank...
Page 84: ...B2AV800 Series Surgical Table IM B2AV800 E01 9 2 This page intentionally left blank...
Page 85: ......