B2AV800 Series Surgical Table
IM-B2AV800-E01
5-20
1. Locate the ratchet wrench (#12, Figure 3-2) under
the right side of Back section, pull down and bring
out the handle.
2. Adjust the lever to the left for the upward
direction, and then use the wrench to bring up the
kidney elevator until it reaches the desired height.
3. Adjust the lever to the right for the downward
direction, and then use the wrench to bring down
the kidney elevator until it reaches the desired
height.
Figure 5-7:
5.2.6 Cassette Channels
There are Cassette Channels beneath the tabletop where radiographs can be taken. Just slide
in the x-ray film cassette (14" wide only).
Figure 5-8:
5.2.7 Use of Equipotentiality for Other Electrical Equipment
WARNING! AVERTISSEMENT!
PATIENT INJURY HAZARD
:
A potential hazard exists when the surgical table is used
together with high-frequency surgical equipment such as cardiac defibrillators and cardiac
defibrillator-monitors. In certain conditions the antistatic properties of the mattress pads
could become
insufficient to prevent burns to the patient’s skin. Care must be taken to
ensure that the insulation between the mattress pad and the patient’s skin is maintained,
including monitoring the bed sheet for excessive moisture, which could compromise the
Lever
Summary of Contents for B2AV800 Series
Page 2: ......
Page 18: ...B2AV800 Series Surgical Table IM B2AV800 E01 1 10 This page intentionally left blank...
Page 32: ...B2AV800 Series Surgical Table IM B2AV800 E01 3 8 This page intentionally left blank...
Page 42: ...B2AV800 Series Surgical Table IM B2AV800 E01 4 10 This page intentionally left blank...
Page 50: ...B2AV800 Series Surgical Table IM B2AV800 E01 5 8...
Page 82: ...B2AV800 Series Surgical Table IM B2AV800 E01 8 2 This page intentionally left blank...
Page 84: ...B2AV800 Series Surgical Table IM B2AV800 E01 9 2 This page intentionally left blank...
Page 85: ......