NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş. 

 
 

NF 048 

 
 

BENCH - TOP CENTRIFUGE 

 

USER’S MANUAL

 

 
 

 
 
 
 
 
 
 
 
 

Z14.K 25 225          Rev. No: 06        Rev. Date: 01/2016 

 

 

Summary of Contents for NF 048

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 048 BENCH TOP CENTRIFUGE USER S MANUAL Z14 K 25 225 Rev No 06 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...PORT 7 4 2 UNPACKING 7 4 3 POSITIONING 8 4 4 MAINS SUPPLY 8 SECTION 5 OPERATING PRINCIPLES 9 5 1 CONTROL PANEL 9 SECTION 6 OPERATING PRINCIPLES 10 6 1 LOADING 10 6 2 PROGRAMMING 11 6 3 MANUAL LID OPEN...

Page 5: ...e microprocessor control system speed RPM or RCF and time could be programmed and the pulse function provides the operator to run the programs of short duration It ensures safe processes by means of t...

Page 6: ...tep 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 60 Hz Power Consumption 450 W External Dim s WXDXH mm 275x355x240 Packing...

Page 7: ...anyone and any hazardous materials should not be in the 300 mm safety zone while centrifuge is running Do not move the device while it is running Do not open the lid while rotor is spinning Apply the...

Page 8: ...the rotor and motor shaft SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol refers to safety relevant warnings and indicates possibly dangerous situation...

Page 9: ...should not be in the 300 mm security area ATTENTION The centrifuges are designed to operate safely under the following conditions Indoor use only Ambient temperature 5 C to 25 C Maximum relative humid...

Page 10: ...also shown on this display 02 Time Display This display shows the time during programming and the run 03 Lid Key and Led This key opens the lid if the led is on It is active before starting the centri...

Page 11: ...ht as the one diametrically opposite for balancing In case the number of tubes which will be centrifuged is less than the capacity of the rotor the tubes must be placed in opposite inserts If the numb...

Page 12: ...kapa anahtar yla bekleme konumuna getirdikten sonra ve yapt n z her program n al mas bittikten sonra yeni bir program al t rmak i in kapa a p kapaman z gerekmektedir 6 3 MANUAL LID OPENING In case of...

Page 13: ...or tube breakage the chamber should be cleaned well Take necessary precautions while cleaning the chamber after tube breakage There could be glass particles in the chamber which may cut the hands Roto...

Page 14: ...entially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for d...

Page 15: ...tor overheat failure Please wait for the motor to cool down and start the centrifuge again Err 6 Motor driver PCB is defective Lid open This failure occurs when lid is opened during the centrifugation...

Page 16: ...15 SECTION 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 NF 048 ELECTRICAL CIRCUIT DIAGRAM...

Page 17: ...16 SECTION 11 WARNING LABEL GROUNDED PLUG...