28
Disinfection can be handled with
NuveDis™
at 90°C whenever it is considered as
necessary by the user.
nUVeRay™
can be used for disinfection.
8.
DISPOSAL MANAGEMENT CONCEPT
The currently valid local regulations governing disposal must be observed. It is in the
responsibility of the user to arrange proper disposal of the individual components.
All parts which may comprise potentially infectious materials have to be disinfected by
suitable validated procedures (autoclaving, chemical treatment) prior to disposal.
Applicable local regulations for disposal have to be carefully observed.
The instruments and electronic accessories (without batteries, power packs etc.) must
be disposed of according to the regulations for the disposal of electronic components.
Batteries, power packs and similar power source have to be dismounted from
electric/electronic parts and disposed of in accordance with applicable local regulations.
9.
TROUBLESHOOTING
If the CO
2
Incubator fails to operate, please check the followings:
The power switch is on;
The fuse is sound;
The plug is plugged-in properly;
The plug is not defective;
The mains supply is present.
In CO
2
alarm conditions, please check the followings:
Check the CO
2
pressure in the CO
2
tank from the pressure regulator.
Check the tube connection between the outlet of pressure regulator and CO
2
inlet on
EC 160 for leakage.
Glass door is closed and the chamber gasket is tight.
9.1.
ERROR CODES
In case of below written failures, related error codes are shown on the display.
Error 1:
Door, chamber and ambient Pt-100 sensor failure.
Error 2:
If the temperature falls down to below 88°C during disinfection, process will
stop and screen goes back to the normal operating mode.
Error 3:
If the temperature exceeds the 95°C during disinfection, process will stop and
screen goes back to the normal operating mode.
Error 4:
CO
2
gas sensor failure.