PERIODIC MAINTENANCE AND CLEANING
7.1.
PERIODIC MAINTENANCE
It is highly reco
mmended to replace the filter inside the chamber once in a year and
replace the filter in the back side of the unit in every 6 months. The period of
changing filter depends on the amount of ambient air pollution and frequency of
usage.
The air circulation pump should be replaced in every 3 years.
nUVeRay™
UV lamp should be replaced once in a year.
Please refer to service manual for further instructions regarding periodic
maintenance.
7.2.
CLEANING
Clean the device when the chamber is at room temperature after disconnecting the
power cable.
Clean the device with a damp cloth to remove dirt and dust.
A soft washcloth must be used not to cause any damage in the chamber.
Do not use alkaline cleaning reagents and metal brushes for cleaning.
Use liquid detergent to remove tough dirt.
Take precautions while handling chemical cleaners. Please be aware of the
undesirable effects of the chemicals and be careful while applying them.
Protect your incubator against rust and ensure any rust spots that may develop are
removed.
The chamber can be cleaned with a surface disinfectant (70% alcohol).
Be careful to remove any encrusted spots of dried medium.
Avoid chlorinated solutions which might damage the chamber.
Shelves, shelf supports and humidity tray can be sterilized at steam sterilizer (20
minutes at 121°C) or at dry heat sterilizer (30 minutes at 180°C).
All operating and error records are
stored in this section. You should be
state date interval for operating
record as section of error record.
Operating and error record data can
be transferred to USB and record
period can be set to given period.