Connect the tube between the CO
2
tank’s regulator and the CO
2
tank port of the
incubator (Figure 2).
Check the connection and be sure there is no leakage.
Turn on the CO
2
tank.
Observe the CO
2
tank’s pressure.
Set the regulator to 2 bars (6-7 liter/minute).
The tube used between CO
2
tank and the instrument should resist to 5 bars.
Tighten the connection of the tube with clamps and ensure that there is no
leakage.
Please prevent the CO
2
gas tank from sun lights and heat. The distance
between tank and a heat source must be at least 0.5 meters. The gas is
inside the CO
2
tank is heavier than the air and has risks for human’s heath.
Figure 7
6.
OPERATING PRINCIPLES
6.1.
SWITCHING ON
Plug-in the instrument to correctly grounded sockets.
Turn on the device by power switch which is at the back of the incubator and see the
LED of power switch is on.
See that the control and the display panel are activated.
Learn the function of the control and display panel (See Section 5.7).
The system controls itself for 60 seconds after it is energized.