11
Indoor use only
A
mbient temperature: 5°C to 40°C
Maximum relative humidity for temperature up to 31
°C: 80%
Maximum altitude: 2000 m
Temperature for m
aximum performance: 15°C / 25°C
5.2.
HANDLING AND TRANSPORTATION
All handling and transportation must be carried out by using proper equipment and
experienced staff. The instrument must be supported underneath and never be turned
upside down.
5.3.
UNPACKING
Remove the cardboard box packing and the second nylon wrapping around the
instrument. Ensure that no damage has occurred during transportation. The below
mentioned are provided with the instrument, please check them;
1 pc. user’s manual and warranty
2 pcs. biological filter
3 pcs. shelves
2 pcs. of shelves’ support
1 pc. shelves’ stick
Humidity tray
1 pc. power cable
2.5 meters tube for CO
2
connection
5.4.
MAINS SUPPLY
The CO
2
Incubator requires 230 V, 50/60 Hz.
Please make sure that the supplied mains matches the required power ratings which are
written on the name of plate of the instrument located at the back of the instrument.
Always plug-in the instrument to correctly grounded sockets.
A supply fitted with a circuit breaker should be used for protection against
indirect contact in case of isolation fault.
5.5.
POSITIONING
Check that the positioning is suitable for the users.
Check that the instrument is stable on its four pedestals.
Check that the user will be able to follow up the operation even when he deals with
something else.
Check that the positioning of the device prevents interference with other equipment
in the near surrounding.
The bench where the instrument is positioned should be resistant to the weight of the
instrument and vibration free.
Leave at least 30 cm space on the left, right and back side of the instrument.