Positioning of EC 160 should be away from windows and walls. The instrument
should be in a place around which the air can ventilate easily.
EC 160 should not be close to the window and exposed to sun light.
Protect the device from direct ventilation.
EC 160 should not be sited on thoroughfares or in line with doorways or opening
windows.
Doors of the room should not be opened frequently.
If there is laminar flow cabinet in the room, it should not blow towards to EC 160.
The air ventilation in the room should be minimum.
If there is air condition in the room, it should be on during night and day. If air
condition is on during day time, do not switch off air condition during night. Or it
should be off all the time.
The room temperature should be stable and it should be maximum 30
⁰
C.
5.6.
GENERAL PRESENTATION
Figure 1
1- C
ONTROL PANEL
2-
CHAMBER GASKET
3-
BIOLOGICAL FILTER
4-
SHELF
5-
GLASS DOOR
6-
UV SOCKET
7-
DOOR SWITCH
8-
HUMIDITY TRAY