NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş.
EC 160
CO
2
INCUBATOR
USER’S MANUAL
Z14.K25 281
Rev. No: 10
Rev. Date: 01/2018
Page 1: ...N VE SANAY MALZEMELER MALAT VE T CARET A EC 160 CO2 INCUBATOR USER S MANUAL Z14 K25 281 Rev No 10 Rev Date 01 2018...
Page 2: ...ate damaged instrument 4 do not operate instrument with a damaged cable 5 do not move instrument during operation In case of a problem contact your N ve agent for an authorized service or maintenance...
Page 3: ...nditions use of the product in non standard environmental conditions including but not limited to failure to meet requirements of ambient temperature lubrication humidity or magnetic field influences...
Page 4: ...ION 10 5 1 ENVIRONMENTAL CONDITIONS 10 5 2 HANDLING AND TRANSPORTATION 11 5 3 UNPACKING 11 5 4 MAINS SUPPLY 11 5 5 POSITIONING 11 5 6 GENERAL PRESENTATION 12 5 7 CONTROL PANEL 15 5 8 PRIOR TO OPERATIO...
Page 5: ...RIODIC MAINTENANCE 27 7 2 CLEANING 27 8 DISPOSAL MANAGEMENT CONCEPT 28 9 TROUBLESHOOTING 28 9 1 ERROR CODES 28 9 2 FUSE REPLACEMENT 29 10 OPTIONS 29 10 1 AlerText GSM ALARM MODULE 29 10 1 1 CONNECTING...
Page 6: ...for reading ambient temperature The volume of EC 160 CO2 Incubator s chamber is 160 liters and is manufactured as in one piece deep drawn stainless steel without any joint in order to prevent contami...
Page 7: ...ch is harmful for the cells The pH of the culture media is kept constant by sodium bicarbonate The chemical reaction of CO2 and sodium bicarbonate can be represented as CO2 H2O H HCO3 2 H CO3 2 The co...
Page 8: ...nsitivity 0 1 CO2 Sensor IR Sensor CO2 Sensor Auto Zero Standard In every 12 minutes Safety System Gas expansion thermostat Chamber Capacity 160 liters Number of Shelves std max 3 8 Power Consumption...
Page 9: ...d be used The set temperature should not destroy the structure of the samples without user s notice The boiling points of the samples should be higher than the set temperature The freezing points of t...
Page 10: ...WA Belgelendirme Hizmetleri A TOSB stanbul Tuzla Organize Sanayi B lgesi Tepe ren Mevkii 34597 Tuzla stanbul TURKEY Symbol in the operating instructions This symbol refers to important circumstances T...
Page 11: ...rs tube for CO2 connection 5 4 MAINS SUPPLY The CO2 Incubator requires 230 V 50 60 Hz Please make sure that the supplied mains matches the required power ratings which are written on the name of plate...
Page 12: ...t be opened frequently If there is laminar flow cabinet in the room it should not blow towards to EC 160 The air ventilation in the room should be minimum If there is air condition in the room it shou...
Page 13: ...13 Figure 2 9 CO2 MEASUREMENT ACCESS PORT 10 AMBIENT TEMPERATURE SENSOR 11 REMOTE ALARM PORT 12 CO2 TANK PORT 13 BIOLOGICAL FILTER 14 SAFETY THERMOSTAT 15 POWER SWITCH 16 POWER SUPPLY AND FUSES...
Page 14: ...O2 Measurement Access Port User could use the port in order to connect an external measurement device and measure the CO2 percentage inside the chamber 10 Ambient Temperature Sensor Measure instant am...
Page 15: ...heir functions correspond to the commands on the display 5 8 PRIOR TO OPERATION 5 8 1 MOUNTING THE SHELVES AND HUMIDITY TRAY Put the shelves supports inside the chamber as it is shown in Figure 4 Figu...
Page 16: ...s place as in Figure 6 Figure 6 Mounting the biological filter outside of the chamber Mount the biological filter into FILTER port located at the back of the instrument as shown in Figure 2 5 8 3 CO2...
Page 17: ...se prevent the CO2 gas tank from sun lights and heat The distance between tank and a heat source must be at least 0 5 meters The gas is inside the CO2 tank is heavier than the air and has risks for hu...
Page 18: ...loaded in order to obtain a uniform temperature distribution 6 2 OPERATING MODE AND MENU 6 2 1 OPERATING MODE NUVE logo appears on the screen for 1 2 minutes for control process This screen is display...
Page 19: ...stribution Door set temperature equals to chamber set temperature value door temperature difference Lock icon appears on the right bottom corner if the device is locked Press that icon to reach the ke...
Page 20: ...rature temperature alarm CO2 CO2 alarm and door temperature difference are shown in this screen Pressing up and down button you can increase or decrease value of parameter after choosing with ENTER F4...
Page 21: ...e and their explanations User orientation aims to make customer usage more interactive and consciously It contains warnings and advices MENU screen are shown by pressing F1 button on operating screen...
Page 22: ...ON selected and approved by pressing the ENTER F4 key warning screens will be seen Take necessary cautions according to warnings There are two types of disinfection process in device These are N veDis...
Page 23: ...omes to screen for informing the user Disinfection operation screen Warning page is passed by pressing F4 button and start option is selected by F2 button and confirmed by F4 button Chamber temperatur...
Page 24: ...t side of the device as in Figure 10 Figure 8 Figure 9 During UV Disinfection process do not to contact with the direct UV light UV lamp should be cleaned frequently against dust and dirt which decrea...
Page 25: ...ushing ENTER F4 key warning screens will be seen Take necessary cautions according to warnings on the screen Use ENTER F4 key to see next screen UV Disinfection time can be set from 15 minutes to 24 h...
Page 26: ...al and it contains device settings In this page user can enter user information into user settings Date and time are selected and entered into date and time section Language section is for language se...
Page 27: ...amage in the chamber Do not use alkaline cleaning reagents and metal brushes for cleaning Use liquid detergent to remove tough dirt Take precautions while handling chemical cleaners Please be aware of...
Page 28: ...tric electronic parts and disposed of in accordance with applicable local regulations 9 TROUBLESHOOTING If the CO2 Incubator fails to operate please check the followings The power switch is on The fus...
Page 29: ...fferent mobile phone numbers in case of failures In addition to this message could be sent as an e mail The data is saved on the internal memory of GSM module in case of failures on the communication...
Page 30: ...NUVE must be informed in advance regarding to SMS message content which will be sent The cost of SIM card and SMS differs according to the GSM network and will be covered by the user 10 2 CO2 CHANGEO...
Page 31: ...al measurement could be done Also an instrument such as magnetic stirrer can be operated inside This port is factory mounted option The port should be sealed well after the cable is passed through aga...
Page 32: ...m unit See Figure 12 Mount the remote alarm kit to a wall or board no more than 10 meters away from the device where it is easily accessible 10 6 nUVeRay UV DISINFECTION KIT The disinfection kit is us...
Page 33: ...33 11 ELECTRICAL CIRCUIT DIAGRAM...