9.2 Electromagnetic immunity
Note:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note:
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz
to
80 MHz
Not applicable Portable and mobile RF communications
equipment should be used no closer to any part of
the THixyz series, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1,2
d = 1,2
80 MHz to 800 MHz
d = 2,3
800 MHz to 2,5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a
should be less than the compliance level in each
frequency range b.
Interference may occur in the vicinity of
equipment marked with the following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz
to
2,5 GHz
3 V/m
Summary of Contents for 10.256.345