6.2.10 Prewarming
On the Refrigerated model, a "Pre-warming" cycle prior to centrifugation allows to warm the centrifuge so it is
ready for a protocol at ambient temperature or warmer (especially after a cycle at a colder temperature).
Visual
Instruction
Control
Access to the list of programs by pressing
this icon :
Select the pre-warming program:
This program is set to warm the centrifuge
up to 20°C during a 20min cycle at 5,100
RPM. Acceleration and Deceleration slope
are both set on 9 and cannot be modified.
Start the cycle by pressing
The temperature inside the bowl will show
on the display
6.2.11 Other menus
Visual
Instruction
Control
Access to other menus by pressing :
Select the menu
•
User Preferences
•
Service menu
•
Information
•
Go back to the centrifugation screen.