OM0252
|
Rev 4 May/2020
Page 22 of 59
NuAire, Inc.
|
2100 Fernbrook Lane
|
Plymouth, MN 55447
|
U.S.A
|
ph: 763.553.1270
|
fx: 763.553.0459
|
tf: 800.328.3352
6.2.7 Interrupt centrifugation
Visual
Instruction
Control
During a centrifugation, press the Start/Stop
button.
The rotor will start slowing down until
complete stop
Open the lid.
6.2.8 Open after an electric power failure / Emergency Stop
This operation must be performed by a person trained in centrifugation-
related risks.
In case of power failure, the lid cannot be opened by the engine unlocking
control.
You must wait 15 minutes for the rotor to come to a complete stop before
opening manually.
1.
Switch off the power to the unit
2.
Insert the unlocking key into the hole on the left side panel
3.
Turn it until the lid opens up
4.
Remove the key
5.
Recover the samples
2
1
3
1
2
3
CAUTION
!