6.2.3 Set program settings
Visual
Instruction
Control
Select the program to edit
Select the desired parameter by pressing
the icon:
Acceleration slope (from 1 to 9)
1: slowest acceleration
9: fastest acceleration
Braking slope )from 0 to 9)
0: no braking at all
1: slowest deceleration
9: fastest deceleration
Centrifuge speed (in RPM or RCF)
Cycle time
Setpoint temperature (only on the
refrigerated model)
Change the value (e.g. speed).
Confirm
Confirm settings
Note: Submit without saving any changes in
the program:
Exit without saving changes: