OM0303
Rev 5 July/2021
Page 87 of 108
o
Blower Window Interlock
This parameter allows for the selection of the window closed switch to be interlocked to the blower.
When the blower interlock is on, the blower will turn off when the window is closed. When the blower
is off, the blower will continue to run when the window is closed.
o
Fan Relay Interlock
This parameter allows for the selection of the fan relay interlock operation. When the fans relay
interlock is on, and the blower switch is pending or blue, the fan relay will be off or not energized. If the
fan relay interlock is off and the blower switch is pending or blue, the fan relay will be on or energized.
In either case the fan relay will be on when the blower switch is on or green and off when the blower
switch is off or not colored.